43
8.8 Initial setting 1
1.
Selection of
Atmospheric Correction
: [ATM CORR]
Select whether Atmospheric Correction is to be performed
by entering the atmospheric temperature and pressure,
by entering ppm value, or by fixing the ppm value to 0
(NIL) not to perform Atmospheric Correction.
2.
Selection of
Prism Constant
: [PRISM CONST.]
Select whether the Prism Constant to be input is set to
0mm, -30mm or to an arbitrary value to be entered from
the keyboard.
3.
Selection for
Reflector sheet Constant
: [SHEET
CONST]
Select whether the target constant to be input is set to
0mm, or to an arbitrary value to be entered from the keyboard.
4.
Selection for
Refraction & Curvature Corrections
: [CRV/REF CORR]
Select whether the correction factor to be input for both
differences (Refraction, Curvature) is set to 0.14, 0.2 or
none (NIL). Selecting “3. NIL” results in no correction of
both values.
5.
Selection of
Tilt Compensation
: [COMP AXIS]
Select whether Tilt Compensation is to be single-axis compensation, dual-axis
compensation or disabled (NIL).
6.
Selection of
Atmospheric corrections display
: [ATM CORR DISP]
Select whether to display the atmospheric corrections in
the title bar of the measure screens or not.
• The factory default for each instrument condition is marked by .
X
1.ATM INPUT
aXXX
2.ppm INPUT
3.NIL
X
1.-30mm
XaXXXXXX
2. 0mm
3.INPUT
X
1. 0mm
XaXXXXXX
2.INPUT
X
1.0.14
XaXXXXXX
2.0.2
3.NIL
X
1.OFF
XaXXXXXX
2.ON
X
1.2 AXIS
XXXXXXX
2.1 AXIS
3.NIL