Safety Instructions
Before lifting
always
ensure that:
• the lift strap is not twisted or worn and can move in and out of the lift unit freely
• the lifting accessories are not damaged
• the sling is correctly and securely applied to the patient in order to prevent injuries from occurring
• the lifting accessory is correctly applied to the lift
• the lifting accessories hang vertically and can move freely
• the sling bar latches are intact; missing or damaged latches must always be replaced
• the sling strap loops are correctly connected to the sling bar hooks when the sling´s strap is extended upward, but before
the patient is lifted from the underlying surface.
Incorrect attachment of sling to slingbar may cause severe injury to the patient.
If the Likorall is installed in the S65 carriage with single hook, make sure it is resting securely at the
bottom of the hook
and is not tilted.
Never leave a patient unattended during a lifting situation!
Only use the Likorall
TM
overhead lift with a carriage, adapter, sling bar and other accessories approved by Hill-Rom.
Before using the lift the first time make sure that:
• the lift is assembled in accordance with the assembly instruction
• the lifting accessories are properly attached to the lift
•
the batteries are charged for at least 8 hours
• you have read the instructions for use for the lift and lifting accessories
•
personnel using the lift are informed of the correct use of the lift and lifting accessories
• the lifting accessory is selected appropriately, in terms of type, size, material and design with regard to the patient’s needs.
No modification of this product is allowed.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the lift, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Use of the product adjacent to other equipment should be avoided because it could result in improper operation, if such use
is necessary, observe and verify that the other equipment is operating normally.
Electromagnetic disturbance, may affect the lifting performance of the product. Modification using other parts than original
spare parts (cables etc.) may affect the electromagnetic compatibility of the product.
Particular care must be observed when using strong sources of electromagnetic disturbance, such as diathermy, etc, so that dia-
thermy cables are not positioned on or near the lift. If you have questions, please consult the responsible assistive-device technician
or the supplier.
The lift may not be used in areas where flammable mixtures may occur, for example in areas where flammable goods are stored.
Installation of Likorall overhead lift to carriages shall be made by personnel authorized by Liko
TM
in accordance with the
installation instructions and recommendations for the current lift system.
Intended use:
The product is intended for use in following environments: Health care, Intensive care, Emergency ward,
Rehabilitation, Habilitation environment. This product is not intended to be used by the patient alone. Lifting and transferring a
patient shall always be performed with the assistance of at least one caregiver. This product is used as a means to perform the lift
but is not in contact with the patient; therefore we do not go into the various patient conditions in this instruction for use. Contact
your Hill-Rom representative for support and advice.
Likorall 200 is tested by an accredited testing institute.