LIGHTMED TruScan Pro
–
Operator’s Manual Rev. No. 01
Page
42
of
109
4.3.3
Pattern Scanner Panel
The pattern scanning function is available for all wavelengths. The controls on the pattern scanner
panel are described in more details below:
Figure 13: Pattern scanner panel controls.
No
.
Item
Description
1
Pattern
display
area
Displays the patterns chosen and/or modified by the user in real time.
2
Pattern
radius
Adjusts the radius between the circular, arc, and curvature patterns. The radius
can be adjusted to 1,000, 1,500, or 2,000 µm.
3
Pattern
spacing
Adjusts the spacing between two dots on any pattern. The maximum spacing
rate is triple (3x) the original size, and the minimum spacing rate would mean
the original pattern spacing.
4
Pattern
rotation
Rotates the pattern by 5° clockwise.
5
Joystick
icon
Inactive
The joystick is inactive.
Activated
The joystick is activated.
6
Single
spot
•
Spot size (µm): 50, 100, 200, 300, 400, 500
•
Power:
o
50 to 500 mW in 10 mW-increments
o
520 to 1,000 mW in 20 mW-increments
o
1,050 mW to maximum power in 50 mW-increments: