Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 88 of 126
C.
Pattern Scan Control Validation
8. Engage the square pattern scanning (5x5) to see if the patterns are evenly
distributed in X and Y axis if not, refer to the service manual realignment
procedures
9. Validate all the control functions in the pattern scan controls screen are
properly functioning
10. Validate if the pattern
D.
Micro Joystick Movement Validation
11. Engage the single spot mode and validate if the joystick able to move in the
field of view, proceed to next for further
If the system rechecking process is valid, then the system is ready to go. More
detailed screen operation and description refer to section 4.2 page xx
E.
Remote Control Verification (Optional)
Prior to commence the final verification, ensure all the corrected setup and
equipment are in placed it.
Accessories:
Truscan Console
Truscan Remote Control
Power cord (110 or 220 VAC type)
Delivery key (Slit lamp / LIO / Endoprobes)
Footswitch (wire or wireless)
Optical fiber (link from console aperture to the LDU could be Endoprobe
or standard delivery one)
Interconnector
On/Off Switch Key
Procedures:
1. Position the laser console onto the left end of side of table top
2. Connect or fit all appropriate connectors including power cord, delivery key,
interlock, footswitch, key switch, and Endoprobe or standard delivery fiber
(refer to fig. 5.19a-b)