![Lightmed LightLas 532/670 Operator'S Manual Download Page 104](http://html1.mh-extra.com/html/lightmed/lightlas-532-670/lightlas-532-670_operators-manual_1912188104.webp)
Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 103 of 126
7.
Ensure the readings fall within the +/-20% tolerance if not, then recheck the
system and make sure the External Meter is set up properly and the fiber has no
sharp bends or damage to it. Replace another fiber to validate it because the
damaged fiber can introduce significant power losses.
8.
If the checking data definitely out of tolerance then the manufacturer or
authorized distributor must be contacted and the Laser System calibration
adjusted according to the procedure in the Service manual.
9.
Be sure to complete the record sheet and keep the copy in the system records.
10.
This system throughput check should be performed at least once a year to
ensure the o/p power performance is within the specified accuracy.
System Throughput Output Power Record Sheet
Power Meter Model:____________ Serial Number:________________
Checked by:__________________________ Date:_________________
Delivery Unit type:____________________________________________
Displayed Power
Meter Reading
Reading X2
Min. / Max.
Pass / Fail
100 mw
80 / 120 mw
500 mw
400 / 600 mw
1000 mw
800/1200 mw
1500 mw
1200/1800mw