Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 – Operators Manual
Rev. No 01
Page 63 of 126
Section 5
Installation
5.1
Introduction
It is strongly recommended that the manufacturer or its authorized agent install the
Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 Pattern Scanning
Laser System at the operator site to ensure that the system is operating correctly,
aligned and calibrated according to specification. After this initial installation it is the
operator’s responsibility to ensure the laser system is operating correctly whenever
the laser is moved or relocated.
The following procedures should be followed in order to successfully install the laser
system. The checklist and report form should be completed and a copy sent to the
manufacturer.
In the event that the report is not sent to the manufacturer then the
manufacturer reserves the right to decline any warranty claims that may be
forthcoming.
The installer should also retain a copy and the customer may keep a copy too.
All precautions care must be taken when installing the Multi-Wavelength Pattern
Scanning System - LightLas 532 / 670 to ensure that you or others are not exposed to
any hazardous laser radiation. Always wear laser safety glasses suitable for the
appropriate wavelength to protect your eyes.
Requirements
1.
100-230 Volts, 50 or 60Hz AC mains power supply with an earth connection.
This is a single-phase outlet capable of delivering up to 400 Watts.
2.
A motorized stand that the Table provided with the Slit lamp can be fitted to
is required if one was not purchased with the Laser system. Ensure the
Table is fitted securely to the Stand using the supplied screws.
3.
A mains power cable is supplied but the connector may not suit the outlet
available so it is advisable to have a spare locally compatible cable available.
The Cable assembly must be CE approved for EU Countries.
4.
A suitable room to place the Laser System in that provides for a safe working
environment is required. As with other ophthalmic equipment a dimly lit
room is preferred.
5.
The Laser System has the facility to connect a remote door interlock to the
Laser treatment room. If this option is required then the customer must
organize this with an electrician and the manufacturer or authorized agent
can provide instructions on how to connect to the Laser System. The Laser
System is provided with a Bypass plug in the event that this option is not
installed. This plug is referred to as the “Interlock Key” and must be