Communications
248083427-1 Rev 4
4-17
Cross-over Cat5 Setup:
1.
Attach one end of the cross-over cable to the RJ45 receptacle on
the PC.
2.
Attach the other end of the cable to the NC30+ Ethernet Port.
3.
Apply power to the NC30+. Apply power to the PC if it is not
already running. Observe the LEDs on the hub/switch RJ45
connector - blinking indicates network activity. Observe the
network LEDs on the instrument’s RJ45 receptacle - they should
blink indicating activity. Proceed to next section.
Program the
Interface
Windows Telnet Programming:
Use Telnet to program the NC30+ for Ethernet communications if
DeviceInstaller is not available.
1.
Start Windows.
2.
On the Taskbar, click on
Start
.
WARNING:
Perform
only the steps or
commands as provided
in this guide. Failure to
heed this warning can
result in damage to
equipment, personal
injury or data loss and
may void the equipment
warranties.
3.
Select
Run
as shown in
Figure 4-12.
Figure 4-12 RUN Screen
4.
In the Run window, type
CMD
and click OK. See Figure 4-13.
Note:
Typing commands
in the command console
requires a space between
the instruction and the
command variables. For
example,’arp’ is followed
by a space, then ’-d’ and
another space, then the *.
Figure 4-13 Starting the Command Console
Summary of Contents for NanoCount 30+
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions NanoCount 30 Liquid Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse NanoCount 30 Operating Manual ii 248083427 1 Rev 4...
Page 70: ...Lighthouse NanoCount 30 Operating Manual 6 6 248083427 1 Rev 4...
Page 86: ...Lighthouse NanoCount 30 Operating Manual B 2 248083427 1 Rev 4...
Page 93: ...00...
Page 94: ......