Step 3: Code the Meter.
While the meter
is on and after the
word
insert
is
displayed, press
and release the C
button; the code
number will
appear. Continue
pressing the C
button until the
number on the
display matches
the code number on the test strip vial.
The code number is now set for the
test strips you are using. The meter
will remember this code number until
you change it. You are now ready to do
a test.
Checking the System
There are two ways to make sure
your OneTouch
®
Profile
®
System
is working properly. The check strip
is used to check that the meter is
operating properly. OneTouch
®
Normal Control Solution is used to
check that both the meter and the
test strips are working together as a
system, and that you are doing the
test correctly. It is very important
that you do these simple checks
routinely for accurate results.