The last Check Strip
reading was outside the
acceptable range of
results.
The Check Strip
test must be repeated.
(See Chapter 3,
Checking the System.
)
If the message
continues, clean the Meter.
(See Chapter
8,
Cleaning and Maintaining the Meter.
)
For further assistance, call the LifeScan
Technical Service Department for more
information.
Usually indicates a test
done with Control
Solution. However, if this
message appears after a
blood test, it means that
your blood sample was
too small, smeared, or a second drop was
applied after the test began.
Your test
result will not be accurate.
Repeat the
test using a larger drop of blood.
(See
Check the Amount of Blood on the Test
Strip in Chapter 6,
Testing Your Blood.
)
All segments on the
display will appear when
the On/Off Button is pressed and held.
If
any part of the display does not appear,
call the LifeScan Technical Service
Department:
Canada
1 800 663-5521
U.S.A.
1 800 227-8862
c 3.8
contrl
Anotok
redo A
•
37
•
(example)
LSI-97-051 OT2.bklt.C.Eg 10/8/97 3:10 PM Page 37 (1,1)