Performance
Characteristics:
See ONE TOUCH
®
Test
Strip package insert for
ONE TOUCH
®
II System
accuracy and precision.
Power Supply:
One Duracell
®
brand
alkaline battery, (
J
-size,
6 volts, part #7K67)
included with Meter. Use
only a Duracell
®
battery
for replacement.
Battery Life:
Approximately one year
(at two tests per day).
Result Range:
0–33.3 mmol/L (0–600
mg/dL). Higher values
displayed as
high
.
Display Type:
Alphanumeric;
6-character,
14-segment LCD.
Dimensions:
Length–12.07 cm (4-3/4").
Width–6.33 cm (2-3/8").
Height–2.7 cm (1-1/16").
Weight–Approx. 135 g
(4.8 oz) (including
battery).
Operating Temperature
Range:
15°C–35°C (59°F–95°F).
Operating Humidity
Range:
0%–90% relative
humidity (non-
condensing).
Memory:
Storage capacity: 250
test results including
Check Strip and Control
Solution results (oldest
test deleted first).
Code Numbers:
1–16.
Data Port:
Communicates with
Data Manager
™
unit and
other data management
systems.
•
84
•
13. SPECIFICATIONS
LSI-97-051 OT2.bklt.C.Eg 10/8/97 3:10 PM Page 84 (1,1)