23
DM00003 Rev D
SYSTEM
MANUAL
5. Plug in the power cord to the back of the photic stimulator, then plug the other end into an electrical outlet.
Fig 11: Photic Stimulator Cable
The Photic Stimulator is classified as a Group 2 instrument under ISO 15004-2. In the event of
an electrical failure that turns the Photic Stimulator on at full brightness and at continuous wave
stimulation (i.e., the photic stimulator turns on and remains on for more than 16 seconds), the light
emitted from this instrument is potentially hazardous. The longer the duration of exposure, the greater
the risk of ocular damage. Exposure to light from this instrument when operated at maximum intensity
will exceed the safety guideline after 36 minutes.
Setting Up the Trackit M