Setting the CH₄ span
Select a quality span gas with a methane concentration near or slightly above the
concentrations you expect to measure.
Caution:
Do not use nitrogen-balanced gas. Use air-balanced gas for both the zero
and span.
1
Power on the instrument and let it warm up.
2
Plumb a gas to the air inlet as shown in
Set the pressure between ambient and 35 kPa (5 PSI) above ambient.
Observe the measurements on the home screen until they have stabilized. The slope
will be near zero.
3
Span the analyzer.
Under
Options > Calibration > CH
4
Span
, enter the concentration of the gas and
click
Span CH
4
.
Section 6. Maintenance
6-3
Setting the CH₄ span