Appendix C. Specifications
Battery Life:
8 hours typical with 2 batteries
Pollution Degree:
2
Over-voltage Category:
II
Class 1 Laser Product
CH
4
Measurements
Response Time (T
10
- T
90
):
≤ 2 seconds from 0 to 2 ppm
Range:
0 to 100 ppm
Precision (1
σ
):
0.60 ppb at 2 ppm with 1 second averaging
0.25 ppb at 2 ppm with 5 second averaging
Maximum Drift:
< 1 ppb per 24-hour period
CO
2
Measurements
Range:
0 to 10,000 ppm
Precision (1
σ
):
3.5 ppm at 400 ppm with 1 second averaging
1.5 ppm at 400 ppm with 5 second averaging
H
2
O Measurements
Range:
0 to 60,000 ppm
Precision (1
σ
):
45 ppm at 10,000 ppm with 1 second averaging
20 ppm at 10,000 ppm with 5 second averaging
C-2