Replacing the air inlet filter
The air intake has filters that may become contaminated periodically. We recom-
mend replacing filters every year, or more often if you are sampling dirty or dusty
air. The filters also should help prevent water from being drawn into the gas ana-
lyzer.
1
Power off the instrument.
2
Remove each of the six screws that secure the air intake and outlet.
They are installed with thread locking compound, which will make them difficult to
remove.
3
Remove the O-rings, the filter support disks, and the filter.
Use a toothpick to work them out.
4
Inspect the O-rings and replace them if they are damaged or compressed.
5
Install the replacement filters and reassemble the system.
Install
two
filters (part number 9978-234) between the support disks. The two metal
discs support the filter(s) — one on each side. Align the centers of the filters and the
metal disks. Assembly is the reverse of disassembly. You do not need to use thread
locking compound on the threads. Tighten them snugly.
Section 6. Maintenance
6-5
Replacing the air inlet filter
Summary of Contents for LI-7810
Page 1: ...LI 7810 CH CO H O Trace Gas Analyzer Instruction Manual ...
Page 2: ......
Page 6: ...iv ...
Page 30: ...Section 2 Getting started 2 14 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 38: ...Section 3 Applications of the LI 7810 3 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 54: ...Section 4 Software overview 4 16 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 88: ...Appendix B Additional safety information B 6 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 92: ...Appendix C Specifications C 4 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 97: ......