Section 6.
Maintenance
The LI-7810 will require little maintenance. There are no user-serviceable parts in
the instrument except for the air inlet filter, air dry-down chemical, and pump. Do
not attempt to disassemble the instrument or conduct any maintenance procedures
other than those described here.
User calibration
To adjust the instrument calibration, you can set the zero and span. The span is like
a slope (the
m
in
y
=
mx
+
b
). The zero is a spectral baseline. Adjusting the zero
changes every measurement (the relationship is non-linear). Adjusting the span
changes the slope. Calibration gases should be at a pressure between ambient and 35
kPa (5 PSI) above ambient pressure. We recommend installing a >30 cm vent before
the air inlet to vent the excess pressure and prevent diffusion of ambient air into the
gas stream.
Figure 6-1. Plumbing configuration for calibration gases. Use a low pressure regulator to
reduce pressure and a 30 cm vent to avoid pressurizing the inlet.
6-1
LI-7810 CH₄/CO₂/H₂O Trace Gas Analyzer
Summary of Contents for LI-7810
Page 1: ...LI 7810 CH CO H O Trace Gas Analyzer Instruction Manual ...
Page 2: ......
Page 6: ...iv ...
Page 30: ...Section 2 Getting started 2 14 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 38: ...Section 3 Applications of the LI 7810 3 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 54: ...Section 4 Software overview 4 16 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 88: ...Appendix B Additional safety information B 6 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 92: ...Appendix C Specifications C 4 LI 7810 CH CO H O Trace Gas Analyzer ...
Page 97: ......