The sections below define several potential problems and lists several possible causes
and their solutions.
Continuous spurious data accumulation
l
Lens cover has not been removed from the camera. Remove lens cover.
l
Camera is at the wrong location. Check the actual camera position as described
in
Changing measurement resolution
l
Fluorescent tube not illuminated. Press "lamp start" switch and hold for 2 or
more seconds.
l
Fluorescent tube is cold. Wait until it warms up.
l
Possible debris within the fluorescent tube. Remove the tube and perform the
procedure in
Voltage selector operation and characteristics
on page 2-5. Altern-
atively, tip the instrument to about 45° by raising the front end to slide the
debris to the end of the tube.
l
Low line voltage. Check it and provide proper voltage to the instrument.
l
Belts are excessively dirty. Clean belts (
Cleaning the transparent belts
l
Mirror, camera, or detector (reticon) need cleaning. Clean according
on page 7-4 and following sections.
l
Obstruction in the optical pathway. Inspect the pathway and remove the obstruc-
tion.
l
Calibration screw adjusted to be excessively sensitive. Note the original position
of the screw and turn the screw counterclockwise until the displayed numbers
remain constant. Return to the original position. Perform calibration as
described in
on page 3-2. If counts continue at an unacceptable rate,
contact LI-COR.
6-1
LI-3100C Area Meter Manual
Summary of Contents for LI-3100C
Page 1: ...LI 3100C Area Meter Instruction Manual...
Page 2: ......
Page 8: ...Section 1 General information 1 2 LI 3100C Area Meter Manual...
Page 14: ...Section 2 Preparations 2 6 LI 3100C Area Meter Manual...
Page 20: ...Section 3 Operation 3 6 LI 3100C Area Meter Manual...
Page 24: ...Section 4 Changing measurement resolution 4 4 LI 3100C Area Meter Manual...
Page 32: ...Section 5 Using the Windows interface software 5 8 LI 3100C Area Meter Manual...
Page 36: ...Section 6 Troubleshooting 6 4 LI 3100C Area Meter Manual...
Page 41: ...brush Section 7 Maintenance 7 5 Cleaning the motor and belt drive system...
Page 42: ...Section 7 Maintenance 7 6 LI 3100C Area Meter Manual...
Page 49: ......