Fluorescent tube installation
Note that the fluorescent tube may be enclosed inside the rear cover of the LI-3100C
during shipping. Remove the rear cover to unpack and install the fluorescent tube.
Figure 2-2. Area Meter in 0.1 mm
2
resolution configuration. (A) front sliding bearing
blocks, (B) "Lamp Start" switch, (C) "Zero" button, (D) sample guides, (E) screws to be
removed in order to detach the front panel.
1
Remove the rear tube connector housing by loosening the two recessed mounting screws (
2
Hold the new tube vertically and gently tap it to cause any internal debris to collect at one end.
Internal debris can cause spurious counting.
3
Slide the tube into the exposed hole in the instrument rear plate. Rotate the tube as needed until it
seats in the front connector. Twist the tube slightly until it snaps into place. You may need to twist
in either direction, depending on the tube's initial position in the front connector.
4
Mount the rear connector housing over the exposed tube end.
5
Gently twist the connector while holding the tube at the back edge of the upper belt. The rotation
should be about 1/4 turn.
Section 2. Preparations
2-3
Fluorescent tube installation
Summary of Contents for LI-3100C
Page 1: ...LI 3100C Area Meter Instruction Manual...
Page 2: ......
Page 8: ...Section 1 General information 1 2 LI 3100C Area Meter Manual...
Page 14: ...Section 2 Preparations 2 6 LI 3100C Area Meter Manual...
Page 20: ...Section 3 Operation 3 6 LI 3100C Area Meter Manual...
Page 24: ...Section 4 Changing measurement resolution 4 4 LI 3100C Area Meter Manual...
Page 32: ...Section 5 Using the Windows interface software 5 8 LI 3100C Area Meter Manual...
Page 36: ...Section 6 Troubleshooting 6 4 LI 3100C Area Meter Manual...
Page 41: ...brush Section 7 Maintenance 7 5 Cleaning the motor and belt drive system...
Page 42: ...Section 7 Maintenance 7 6 LI 3100C Area Meter Manual...
Page 49: ......