Plug-in control
37
GA 05.139/1.02 - 06/99
Start
Stop
Esc
Mon
Prog
Enter
+
_
Acceleration
17.5A 254Hz HOK
Key
Function
Esc
• Returns to the operating display from the storage procedure
without storage.
• Returns to the operating display from any point of the basic
menu.
Mon
• No function
Prog
• Selects the programming menu from the operating display.
• Confirms to store changed parameters to the EEPROM
Enter
• Switches forward to the next submenu
Start
• Starts the pump (only possible if there is no fault).
The start key is only active if the user is in the basic menu
or in the operating display.
Stop
• Stops the pump
• Returns to the operating display from the programming menu.
• Acknowledges a failure after the cause of the failure has
been removed.
+
• Increases a parameter value or proceeds to the next option.
–
• Lowers a parameter value or returns to the previous option.
<
• Selects the programming menu from the operating display.
• Switches back to the last main menu.
>
• Selects the programming menu from the operating display.
• Switches forward to the next main menu.
Fig. 33 Functions of the front panel keys
Acceleration
17.5A 254Hz HOK
Operating display
Operating status
Motor current / Actual frequency / TMS status code
If a critical operating status occurs this warning is
displayed alternating with the operating display.
Failure
Accel. Time
Failure message
Failure cause
TMS status code
No.
Code Description
——————————————————————————
1
H
Heating pump
2
H O K
Temp. TMS ok, heater ON
3
O K
Temp. TMS ok (range T
set
± 2K)
4
C O K
Temp. TMS ok, cooling ON
5
W
Warning Temp. TMS
O F F
TMS cancelled via control plug X14
see also Section 4.4 „TMS“
Fig. 34 Display