1-point teach of the analog output

First activate the previously described teach mode for output Analog OUT.

By selecting an object distance within the scanning range, the characteristic curve of the analog output can be adjusted.

If an object is located outside of the taught measurement range, an error signal is output. A different analog signal is output here by the 
sensor for the errors "distance too close: object outside of the measurement range" and "distance too far: object outside of the 
measurement range".

2-point teach of the analog output

First activate the previously described teach mode for output Analog OUT.

By selecting 2 object distances within the scanning range, the characteristic curve of the analog output can be adjusted. 

If an object is located outside of the taught measurement range, an error signal is output. A different analog signal is output here by the 
sensor for the errors "distance too close: object outside of the measurement range" and "distance too far: object outside of the 
measurement range".

Inverting the analog output (falling/rising characteristic curve)

First activate the previously described teach mode for output Analog OUT.

The characteristic curve of the analog output can be inverted, e.g., if a falling characteristic output curve is desired. 

1-point teach - rising characteristic curve 

1. Place

 object at desired distance for the end point of the measurement range.

Note

:

The 

minimum object distance for the end of the measurement range

 is as follows:

scanning range of 400mm:

90mm

scanning range of 1600mm:

310mm

2.

To adjust analog output 

Analog OUT

press

 the 

teach button

 for 

2 … 7s

 

until the 

blue and green LEDs flash simultaneously at 3Hz

.

3. Release the button

. The characteristic curve with plot rising from the start of the range (50 mm or 150 mm) to the set object distance was taught in. 

4.

Error-free teach: LED states acc. to "Technical data" -> "Indicators". 

Faulty teach: green and blue LEDs flash at 8Hz

 until an error-free teach is performed.

2-point teach - rising characteristic curve 

1. Position

 the object at the first desired distance (near or far).

2.

To adjust analog output 

Analog OUT

press

 the 

teach button

 for 

7 … 12s

 until the 

blue and green LEDs flash alternately at 3Hz

.

3. Release the button

. The sensor remains in teach mode and the LEDs continue to flash. 

4.

Then 

position

 the object at the second desired distance (far or near). 

Note

:

the 

minimum object distance between the start and end point of the measurement range

for a scanning range of 400mm is:

40mm

for a scanning range of 1600mm is:

160mm

5. Briefly press

 the 

teach button

 again to complete the teach event. 

The characteristic curve with rising plot from the near to the far object distance was taught in. 

6.

Error-free teach: LED states acc. to "Technical data" -> "Indicators". 

Faulty teach: green and blue LEDs flash at 8Hz

 until an error-free teach is performed.

Inverting the characteristic curve 

1.

To invert the characteristic curve of the analog output 

Analog OUT

press

 the 

teach button

 for 

> 12s

 until the 

blue and green LEDs flash alternately

.

2. Release the button

. The characteristic curve plot was inverted.

The 

blue LED

 indicates the current setting of the analog output: 

ON

rising

 characteristic curve

OFF

falling

 characteristic curve

DMU318…W

Summary of Contents for DMU318 W Series

Page 1: ...PN NO NC function reversible NEW Both outputs can easily be taught using a button NEW Stable plastic design NEW Temperature compensated scanning range 50 400mm 150 1600mm 10 30 V DC 4 20mA 0 10V Acces...

Page 2: ...utton 12s Indicators Yellow LED OUT2 object detected Blue LED Analog OUT object detected Yellow green or blue green LED flashing Teach in teaching error Green LED Object within the scanning range Mech...

Page 3: ...contact preset C Analog output 4 20mA V Analog output 0 10V Pin assignment of connector pin 5 gray cable wire Sync MUX K Synchronization multiplex input Connection technology M12 M12 connector 5 pin...

Page 4: ...rmally closed Close Far Far Close Dead zone Sensor OUT1 active Yellow LED ON Green LED OFF OUT1 not active Yellow LED OFF Green LED ON Distance Hysteresis Switching point Scanning range OUT1 not activ...

Page 5: ...th 2 point window teach as a function of the switching function 1 Place object at desired switching distance 1 First place object at desired switching distance for switching point 1 2 To adjust the ou...

Page 6: ...output is changed to the opposite state toggled First activate the previously described teach mode for output OUT 1 Changeover of the switching function 1 To change the switching function of output O...

Page 7: ...that is to be taught OUT1 or Analog OUT 1 Press the teach button for 2s to activate teach mode The yellow LED OUT 1 flashes at 1Hz While in this state output OUT 1 can be taught 2 To teach output Ana...

Page 8: ...ning range of 400mm 90mm scanning range of 1600mm 310mm 2 To adjust analog output Analog OUT press the teach button for 2 7s until the blue and green LEDs flash simultaneously at 3Hz 3 Release the but...

Page 9: ...of the same type are wired together in a network according to the following diagram The devices work in synchronous operation with a simultaneous transmission pulse The response time of the individual...

Page 10: ...factory setting 1 switching point at 400mm or 1600mm rising characteristic curve with spread over the entire scanning range Resetting to factory settings 1 When switching on the supply voltage during...

Reviews: