Entering further basic program data
29
Error messages
All error messages are lost after a power failure.
Certain instrument faults always trigger an alarm or a message
(see chapter “Description of the relay outputs”)!
Function AUX
The AUX relay output can be programmed for the following control
functions:
– As a function output for the contact with programmable duration
prior to and/or during the analysis, or after an analysis.
For example, it is possible to control the cooling water inlet of an
upstream cooler via a solenoid valve. This ensures that the cooling
water only flows when required, i.e. when an analysis is being carried
out.
In >
, select
=>
=>
Select the program step at which the AUX contact is to be
activated
Under "Time" enter the contact duration in minutes (m) and
seconds (s)
Press "ENTER" to confirm the entry
▼▲
NOTE