Program menu "M"
40
Program menu "M"
Call:
(1)
Use the key to open the program menu "M".
It is possible to call up all the functions without password protection
except for the basic program.
Programming of: reagents, manual mode, flushing, flush chamber,
drain chamber, fill chamber, self-test, confirm maintenance,
diagnosis, date, time, basic program with password
Availability of functions
All manual functions can only be selected during an analysis
pause. Analyses are not carried out during manual operation. All
signal inputs and outputs are locked.
Fill chamber (8)
Press the "ENTER" key to fill
the
measuring chamber.
Flush chamber (6)
Press the "ENTER" key to flush
the measuring chamber once.
Drain chamber (7)
Press the "ENTER" key to open
the outlet valve in order to drain
the water from the measuring
chamber. Press the "ENTER"
key again to cancel this
function.
Service I (2)
Input reagent (3)
Enter the new filling levels after
each
refill or reagent bottle
change. Once you have
selected the menu item for
entering the filling level
"Reagent X Filling (0 - 100%)"
via the "ENTER" key, the value
is preset to 100%. If a full bottle
has been connected, confirm
this value via the "ENTER" key.
If the filling of the bottle differs,
enter the corresponding value.
Manual operation (4)
After confirming the information
message (4) via the "ENTER"
key, it is possible to select and
activate the desired function by
using the arrow keys and
pressing the "ENTER" key.
These functions are used for
checking the functions and for
commissioning.
Flush (5)
Start the flushing of the
sampling tube through the
internal valve by pressing
"ENTER". When "ENTER" is
pressed again, this function is
terminated.
NOTE