Important safety information
6
Installation
Always completely disconnect the relevant plant part before in-
stalling the instrument or connecting/disconnecting it to/from the
power supply. Secure the plant against reconnection.
Only connect the instrument to the mains voltage specified on the
rating plate.
Always observe technical data and ambient parameters.
Testomat 2000
®
Fe requires an interference free and stable power
supply. If necessary, use a mains filter to protect Testomat 2000
®
Fe against interference voltages caused, e.g., by solenoid valves
or large motors. Never lay connecting cables parallel to power ca-
bles.
Operation
Ensure that the maximum electrical load capacity of the relay out-
puts is never exceeded.
Immediately switch off Testomat 2000
®
Fe and contact service
staff if malfunctioning occurs. The warranty will be void if you tam-
per with or attempt to repair Testomat 2000
®
Fe. Repairs must be
carried out by authorised service staff.
After switch-off and longer downtime
Ventilate the indicator leads as described in the start-up, because
longer downtimes (more than 6 hours) can cause the indicator in
the leads to retract.
Do not switch off the appliance for longer periods (e.g. over the
weekend) via the start/stop output. The indicator can retract from
the leads. This results in measurement errors after the appliance is
switched on.
Cleaning
Only use a dry, lint-free cloth for cleaning.
De-installation
Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a
defective instrument (irrespective of the warranty period) if it has
been de-installed and returned to us with a description of the error.
Please also inform us of the measured medium. Before you return
the instrument for repair work, remove the reagent bottles and en-
sure that the measuring chamber has been flushed out and is
empty.
Disposal
Dispose of the instrument in accordance with national regulations.
WARNING
!
NOTE