Maintenance
47
Maintenance
Required maintenance measures
Regular maintenance is necessary to ensure trouble-free opera-
tion of the instrument!
Please regularly carry out the maintenance work described in the
following section when
the programmed maintenance date has been reached (display
"maintenance date exceeded")
the instrument displays the following error messages:
"Mf. soiling" or "Low reagent level"
the last maintenance was carried out max. 6 months ago
Cleaning measures
Never use organic solvents to clean the measuring chamber or
other plastic parts!
Please observe the safety regulations when handling cleaning
agents!
If the measuring range of the instrument is exceeded over a long-
er period of time, a coloured film may form on the sight-glass win-
dows. Use alcohol to remove this sticky film.
Description of maintenance work
Please refer to the "Maintenance manual Testomat 2000
®
/Testomat
ECO
®"
for a detailed description of maintenance work. The measures
described here only provide a brief overview.
Cleaning the measuring chamber and sight-glass windows
Switch off the instrument or press the "STANDBY" key. If required,
remove any water from the measuring chamber.
Close the manually-operated valve of the branch line to Testomat
2000
®
Fe.
Unhook the toggle type fastener
, tilt the measuring chamber
upwards and remove it.
Release both sight-glass window holders
; remove and clean
the sight-glass windows.
Use alcohol to clean off the film on the sight-glass windows. If the
instrument has been used with hard water for a longer period of
time, a hard-to-remove film may have formed on the sight-glass
windows. If this is the case, clean the sight-glass windows as de-
scribed below for cleaning the measuring chamber.
CAUTION
!
NOTE