Program menu "M"
42
Program menu "M"
Call:
(1)
Use the key to open the program menu "M".
It is possible to call up all the functions without password protection
except for the basic program.
Programming of: reagents, manual mode, flushing, flush chamber,
drain chamber, fill chamber, self-test, confirm maintenance, diagno-
sis, date, time, basic program with password
Availability of functions
All manual functions can only be selected during an analysis
pause. Analyses are not carried out during manual mode. All sig-
nal inputs and outputs are locked.
Service I (2)
Input reagents (3)
Enter the new filling levels after
each
refill or reagent bottle
change. Once you have select-
ed the menu item for entering
the filling level "Reagent A
filling level (0 - 100%)" or "Re-
agent B filling level (0 - 100%)"
via the "ENTER" key, the value
is preset to 100%. If you have
connected a full bottle, press
"ENTER" to confirm the value.
If the filling level of the bottle
differs, enter the corresponding
value.
Manual operation (4)
After confirming the information
message (4) via the "ENTER"
key, it is possible to select and
activate the desired function by
using the arrow keys and
pressing the "ENTER" key.
These functions are used for
checking the functions and for
commissioning.
Fill chamber (8)
Press the "ENTER" key to fill
the
measuring chamber.
Flush (5)
Start the flushing of the sam-
pling tube through the internal
valve by pressing "ENTER".
Press "ENTER" again to termi-
nate this function.
Flush chamber (6)
Press the "ENTER" key to flush
the measuring chamber once.
Drain chamber (7)
Press the "ENTER" key to open
the outlet valve in order to drain
the water from the measuring
chamber. Press the "ENTER"
key again to cancel this func-
tion.
NOTE