Entering further basic program data
26
Interval pause
If the analysis is started via a timer, the interval between two anal-
yses (plus flushing time) is determined by the interval pause. The
shortest interval can be 0 minutes. In this case, analyses are carried
out continuously. The longest interval is 99 minutes.
In the
menu, select
=>
=>
Enter the "
in minutes (m)
Press "ENTER" to confirm the entry
Limit value monitoring
It is possible to program the limit values on a continuous scale. The
limit value range depends on the programmed unit. Two limit value
outputs are available for monitoring. This ensures that two limit val-
ues or two measuring points can be monitored. The functions of the
allocated relay outputs can be programmed independently of each
other.
If the unit is used for monitoring two
limit values
, the limit value out-
puts are permanently allocated to these limit values!
1
LV1 = Limit value 1
2
LV2 = Limit value 2
If the unit is used for monitoring two
measuring points
, the limit value
outputs are permanently allocated to these measuring points!
1
LV1 = Measuring point 1
2
LV2 = Measuring point
2
If the limit value LV1 has been exceeded, the limit value control dis-
play
1
lights up RED
and the relay output LV1 reacts as pro-
grammed in the switch function. If the limit value has not been ex-
ceeded, the display lights up GREEN
.
The same applies for the limit
value LV2.
In the
menu, select
=>
=>
Enter the values for "
or "
Press "ENTER" to confirm the entry
Monitoring of two
limit values
Monitoring of two
measuring points
▼▲
▼▲