• Speed:
Test speed in mm / min or inch /min (depending on the settings below)
• Initial Length: Clamping distance in mm or Inch of the device jaws after actuation with the key
–
‹‹
oder
+
››
.
The jaw spacing can be corrected at any time with the key
–
‹‹
or
+
››
.
The absolute position of the
sliding carriage (12)
(jaw spacing) appears on the
display (4)
.
• Initial Tension: Level of initial tension. When the set initial tension is achieved, the evaluation of the tensile
test starts. If the initial tension is set to
O
, the test evaluation can be started by pressing the
›
Start key.
• Exit menu by pushing
SET
key
• When setting the initial length, the force display may show the value ≠ 0. Reason: Influence of temperature
on the equipment and/or force on the
clamping jaws (11)
which are close to the
housing of the drive
motor and circuit board (15).
• When starting the test, the force values are reset to 0.
Version with USB port has more menu items (see operation of USB port).
• Open storage case
• Connect enclosed power
cable to socket (1)
.
• Connect the machine to the line/mains
• Pull out
clamping lever (9)
un till it is locked
• Turn-on main
switch (2)
– Position of sliding carriage is not at
spacer (13)
– Press <<, for Initialize appears on
display (4)
– Push key
–
‹‹
, Wait for Initialize appears on
display (4)
and
sliding carriage (12)
moves to
spacer (13)
.
–
Display (4)
shows standard mode
– Position of sliding carriage is at
spacer (13)
–
Display (4)
shows standard mode
• Set the test parameters using the following keys:
Readiness for working
Test parameter
Display unit
SET
–
‹‹
+
››
mm inch
Display unit
SET
–
‹‹
+
››
N
lbf
Set Initial Tension
SET
–
‹‹
+
››
Set Initial Length
SET
–
‹‹
+
››
Set speed
SET
–
‹‹
+
››