background image

11

6.3.  Test with VACUUM PLATE

6.2.  Preparation and operation

Turn clockwise (+)

pressure setting valve closed/maximum pressure

Turn anti-clockwise (-)

pressure setting valve open/minimum pressure

•  The welding area must be dry and free of contamination. Dirt particles may compromise the test.
•  The seam area to be inspected must be sprayed with a liquid forming bubbles (e.g. soapy water) directly 

before the inspection. The liquid must not damage the seam and the weld material. 

• 

Set the pressure setting valve (Vari-Flow Valve) (9)

 to minimum pressure before the test. Turn counter-

clockwise until stop (-). 

•  Position the VACUUM PLATE on the location to be tested 
•  Place the VACUUM PLATE so that the seam to be tested is approximately in the middle of the longitudinal axis 

of the testing device. For longer test sections, the VACUUM PLATE needs to be repositioned accordingly. When 
moving the VACUUM PLATE, the test areas must be overlapped.

•  In order to generate a negative pressure, the VACUUM PLATE must be pressed lightly on the location to be 

tested with both

handles (8).

• 

Turn on the main switch (2)

 on the VACUUM PLATE

•  Turn the 

pressure setting valve (Vari-Flow Valve) (9)

 clockwise (+) until the desired or specified test 

pressure is set.

•  The existing negative pressure can be read on the 

manometer (5)

.

•  Visual inspection of the weld seam for leaks
•  The seam is sealed if there are no bubbles in the test liquid.
• 

Switch off the main switch (2)

 of the VACUUM PLATE, let the device cool down 

•  Then disconnect the 

power cord (1)

 from the electrical mains. 

•  Mark and repairany leaks

•  Before each start, check the VACUUM PLATE for faults and damage, e.g. cracks, splits or scratches; a defec-

tive VACUUM PLATE can implode

•  Visually inspect that the 

filter plate (4)

 is present and whether it is clean. 

• 

Check the power supply line and plug (1)

 as well as the extension cable for electrical and mechanical 

damage. 

• 

Check the cooling air openings (11)

 must be open at all times and air must be able to be suctioned and 

purged unhindered 

•  Ensure the 

sealing lip (7)

 is not damaged 

• 

Warning: 

Excessive damage to the sealing

 lip (7)

 may prevent the required test pressure from being reached.

•  Connect the VACUUM PLATE

 to the nominal voltage

. The 

nominal voltage

 specified on the device must 

match the 

supply voltage

.

Summary of Contents for 169.580

Page 1: ...Leister Technologies AG Galileo Strasse 10 6056 Kaegiswil Switzerland Tel 41 41 662 74 74 Fax 41 41 662 74 16 www leister com sales leister com VACUUM PLATE 300 English...

Page 2: ...7 5 3 Overview of device parts 8 6 Operation of your VACUUM PLATE 300 9 6 1 Work environment and safety 9 6 2 Preparation and operation 11 6 3 Test with VACUUM PLATE 11 7 Quick guide VACUUM PLATE 300...

Page 3: ...vapor and grease must not be sucked in Any operation in a potentially explosive environment is prohibited The device must only be operated with the filter plate installed No hazardous leak tracer is t...

Page 4: ...are of hazards in the surrounding area e g risk of tripping risk of slipping strong sunlight unattended equipment etc Note Caution 2 2 Non intended use 2 1 Intended use Any other use or any use beyond...

Page 5: ...Ambient temperature C F 10 60 14 140 Noise level LA LPA dB LWA dB 84 7 K 3 dB 95 7 K 3 dB Test pressure bar inHg 0 170 5 Weight kg lbs 7 7 17 a b c Dimensions a mm inches 750 29 5 b mm inches 250 9 8...

Page 6: ...mechanical damage The device must always be stored on a flat surface otherwise the sealing lip 7 can become permanently deformed Note Do not use the carrying handle 8 on the device for transporting wi...

Page 7: ...event of any inquiries to our representatives or authorized Leister Service Centers please always refer to this information Model Serial no 5 2 Scope of delivery Standard equipment in cardboard box 5...

Page 8: ...ressure setting valve Vari Flow valve 10 Type plate with model designation and series marking 11 Cooling openings 12 Suction air outlet 13 Device protective circuit breaker 1 Mains connection line 2 M...

Page 9: ...ial safety data sheet from the manufacturer of the material and follow that company s instructions Risk of injury from plastic chips at startup Damaged plastic glass can lead to implosion This may cau...

Page 10: ...g risk of tripping risk of slipping strong sunlight unattended equipment etc Note Caution 230 V up to 50 m 3 1 5 mm2 up to 100 m 3 2 5 mm2 120 V up to 50 m 3 1 5 mm2 up to 100 m 3 2 5 mm2 Power cord...

Page 11: ...tested with bothhandles 8 Turn on the main switch 2 on the VACUUM PLATE Turn the pressure setting valve Vari Flow Valve 9 clockwise until the desired or specified test pressure is set The existing ne...

Page 12: ...power supply cord 1 to the voltage source 2 Open the pressure setting valve Vari Flow Valve 9 turn counter clockwise to the stop 3 hold the device by the carrying handles 8 place it on the test locati...

Page 13: ...Flow Valve 9 Check the flatness of the membrane while pressing the VACUUM PLATE The VACUUM PLATE can compensate an unevenness of approx 10 mm via the sealing lip 7 There is a risk of leakage if there...

Page 14: ...ntact with aggressive substances e g acetone diluent etc There is a risk of damage to the installed components Recommended maintenance intervals for the device Note Dangerous voltage danger to life fr...

Page 15: ...65 EU Harmonized EN ISO 12100 2010 EN 60335 1 2012 A11 2014 A13 2017 AC 2014 Standards EN 55014 1 2017 A11 2020 EN 55014 2 1997 A1 2001 A2 2008 AC 1997 EN IEC 63000 2018 Kaegiswil 02 24 2022 Bruno von...

Page 16: ...y claim verification by invoice or delivery note manufacturing or processing errors will be rectified by the sales partner through replacement delivery or repair Heating elements are excluded from war...

Reviews: