8.2.2 Maintenance instructions
Only authorized and qualified service personnel may access the internal components of the
instrument for service and repair!
Note for inspection
•
When the instrument reaches approx. 2 million performed strokes, the
code
SEr
always appears for approx. 3 seconds in the three-digit
display when the
TRIM/SECT
button is pressed
(switching between trimming and sectioning mode).
This is a reminder that the instrument must be inspected by an autho-
rized Leica service technician, regardless of how heavily the instru-
ment has been used.
↓↓↓↓↓
The instrument is basically maintenance-free.
To ensure trouble-free operation of the instrument over a long period of time, the following is
recommended by Leica:
3.
Have the instrument inspected on a regular
basis by a qualified service technician au-
thorized by Leica.
The intervals depend on how heavily the
instrument is used.
1.
Thoroughly clean the instrument on a daily
basis.
2.
Enter into a service contract at the latest at
the end of the warranty period. For more
information, please contact your local Leica
technical service center.
We recommend the following inspection intervals depending on the general workload of the instru-
ment as defined in the two categories listed in the table:
8.
Cleaning and maintenance
Category I
> 8.000 sections per day
> 5 hours per day
Predominantly high sectioning speed
Working with soft and hard specimens
Every 12 months
Number of sections per day:
Workload (hours per day):
Sectioning speed:
Specimen material:
Maintenance:
Category II
< 8.000 sections per day
<
5 hours per day
Slow to medium sectioning speed
Predominantly soft specimen material
Every 24 months