19
Leica RM2255
4.
Startup
22
25
24
21
23
The voltage selector is located above the main
power switch, on the left-rear side of the instru-
ment (
Fig. 12
). The voltage setting is displayed in
the viewing window (
22
).
•
Insert a small screwdriver into the cutout
(
25
) and carefully pry out the insert.
•
Remove the voltage selector housing (
21
)
together with the fuses (
23
). Remove the
voltage selector block (
24
) (white) and insert
it again such that the correct local voltage is
displayed in the viewing window (
22
).
•
Reinsert the voltage selector housing with
the block and fuses and push it in until it
engages (audible click).
21
26
Connecting the power supply
•
Before connecting the power cable, make sure that the power switch
(
27
) on the rear of the instrument is switched to "
O
" =
OFF
.
•
Various country-specific power cables are provided with the instru-
ment. Make sure that the power cable used has the correct plug for
the power socket.
•
Insert the connector of the power cable into the connection socket
(
26
) and plug the power plug into the power socket.
Exposure to extreme temperature changes and high air humidity
may cause condensation to form inside the instrument.
After transporting, please wait at least 2 hours to allow the
instrument to adopt the ambient temperature before turning it on!
Failure to comply with this may cause damage to the instrument.
27
Fig. 12
Checking the voltage
(continued)