42
Instruction manual V 1.6 – 10/2009
6.
Optional accessories
Clamping of foil pieces
•
Move the movable jaw (
74
) to the right as
required by turning the setscrew with an
Allen key size 4 (
71
).
•
Place the foil (
72
) between the movable jaw
(
74
) and the fixed jaw (
73
).
•
To clamp the foil, screw the movable jaw
(
74
) against the fixed jaw (
73
) by using the
Allen key.
•
Insert the foil clamp (
75
) in the standard
specimen clamp as shown.
•
Turn the knurled screw (
66
) clockwise to
clamp the foil clamp in the standard speci-
men clamp.
6.2.3 Foil clamp type 1
The foil clamp type 1 is appropriate both for
clamping very small foil pieces and flat, angular
specimens. It is mounted in the standard speci-
men clamp.
Clamping of flat, angular specimens
To clamp angular specimens, replace the long
setscrew (
76
) with the short setscrew (
77
) pro-
vided with the foil clamp.
•
Unscrew the long setscrew (
76
) to the left
with an Allen key size 4 (
71
).
•
Screw the short setscrew (
77
) in the hole.
•
Place the specimen (
67
) between the
movable jaw (
74
) and the fixed jaw (
73
).
•
To clamp the specimen, screw the movable
jaw (
74
) by turning the setscrew (
77
) against
the fixed jaw (
73
).
•
Insert the foil clamp in the standard speci-
men clamp as shown.
•
Turn the knurled screw (
66
) clockwise
to clamp the foil clamp in the standard spec-
imen clamp.
Fig. 35
Fig. 34
66
74
72
75
71
73
76
77
67
74
73