Instruction manual V 1.6 – 10/2009
5.
Operation
•
To change the display mode, push
MENU MODE
until the LED of the
desired mode is illuminated.
•
Press
CLEAR
to reset section thickness sum or section number.
•
This will only reset the currently displayed value.
Caution!
When the instrument is switched off using the main power switch,
both values (section thickness sum and section number) are
erased from memory.
Four-digit display on the instrument
The four-digit display is adjustable.
When the
ΣΣΣΣΣ
µm
LED is lit, the display shows the
sum of the section thicknesses in µm for all sec-
tions completed since the instrument was
switched on.
(Section thickness sum)
When the
ΣΣΣΣΣ
n
LED is lit, the display shows the
number of all previously completed sections.
Fig. 17
•
Press the button to start the slow backwards movement. The
travel continues as long as the button is held depressed.
•
Press the appropriate button to start a rapid or slow forward movement.
The travel continues as long as the button is held depressed.
•
During the forwards movement, the yellow LED (
41
) in the button flash-
es. When the front end position is reached, an acoustic signal is heard
and the LED stops flashing and remains lit.
41
Forward coarse feed
Specimen retraction
To prevent damage to the blade and specimen, the specimen is moved
away from the blade during the return motion to the upper home position.
The retraction value can be selected in 5 µm increments between 5 and
100 µm. Specimen retraction is set to 10 µm at the factory.
Specimen retraction can be disabled if required.
The selected setting is maintained when the instrument is turned off.