7.
Adjustment Accuracy
After pressing the last
in the wizard the results are shown and
stored to the instrument.
8.
Tap to get back to the
Devices
page.
5.4
Adjusting the Circular Level of the Instrument and Tribrach
16549_001
☞
A field controller is necessary for this procedure.
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Turn on the instrument. Connect the instrument to the field control-
ler. Using the tribrach footscrews, level the instrument with the elec-
tronic level.
☞
Electronic level:
•
Select
Devices
from the Home Menu.
•
Tap the arrow button to the right of the device name.
•
To display the electronic level, tap
Compensator
.
3.
Check the position of the circular level on the instrument and tri-
brach.
☞
If both circular levels are centred, no adjustments are necessary.
Adjusting the circular
level step-by-step
48
Check & Adjust