5. Assembling the microscope – trAnsmitted light illuminAtion cArrier
5.8 Assembling the transmitted light illuminator arm (TL)
5.8.1 Tiltable transmitted light illuminator arm
• Wipe off the microscope’s contact surface (Fig. 26.3) with
a dry cloth.
• Lightly tilt the illuminator arm (Fig. 26.1) to the rear and
insert it so that the pin (Fig. 26.2) fits in the groove of the
contact surface (Fig. 26.4).
• Raise up the transmitted light illuminator arm and fasten it
with 4 screws (Fig. 27.1).
When fastening, do not hold onto the transmitted light il-
luminator arm. This ensures optimum alignment for the
optical axis.
• The tilting angle of the illuminator arm can be modified or
locked in place in the vertical position using the knurled
screw (Fig. 27.2).
Leica DMi8 A
• Connect the electronics cable to one of the EXT1–EXT6
sockets (Fig. 27.3).
Fig. 26:
Transmitted light illuminator arm
assembly
1
Transmitted light illuminator arm
2
Pin on transmitted light illuminator
arm
3
Contact surface
4
Groove on contact surface
5
Connecting cable
1
3
4
5
2
Fig. 27: Rear side of motorized Leica DMi8
1
Fastening screws on the transmitted light illuminator arm
2
Knurled ring for locking the transmitted light llluminator arm in
place
3
EXT1-EXT6 sockets
1 2
3
1