6. Assembling
Fig. 16 a-c
Lamp mounts for
gas discharge lamps
1
Upper clamping system,
2
Lower clamping system,
3
Cooling element,
4
Melt nipple for the Hg 50 arc lamp,
5
Dust cover for the Xe 75 arc lamp
a
b
3
Hg 50
Xe 75
1
4
2
3
1
2
5
• To open the 106z lamp housing, unscrew the
fastening screws (15.8) on the cover.
• Remove the transport anchor (red plastic rod
in place of the burner) in the lamp mount. To do
so, remove the lower clamp (16.1). Pull up the
cooling element (16.3) and turn it to the side.
Detach the lower clamp system (16.2) and re-
move the transport anchor.
• Install the burner in reverse order.
Caution!
Hg 50 burner:
After installation, the labeling must be up-
right.
If a glass melt nipple is present (16a.4), posi-
tion it by turning the burner so that the nipple
does not impede the beam path later, but in-
stead is positioned sideways.
Xe 75 burner:
Remove the burner’s dust cover (16b.5) after
you have installed the burner.
c
Hg 100
3
1
2
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (15.8).
• Close the lamp housing and retighten the fas-
tening screws.
• Place the lamp housing in the incident light
lamp housing receptacle (11.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the external pow-
er supply.
Summary of Contents for DM2700 M
Page 1: ...Leica DM2700M Instructions Bedienungsanleitung ...
Page 3: ...Leica DM2700M Instructions ...
Page 18: ...18 4 Overview of the Instrument 410 496 330 115 Dimensions specification in mm ...
Page 79: ...Leica DM2700M Bedienungsanleitung ...
Page 94: ...18 4 Geräteübersicht Abmessungen Angabe in mm 410 496 330 115 ...
Page 155: ......