9.3.2 Replacing the fuses
9.
Cleaning, Disinfection, Maintenance
On the back of the instrument there is a fuse box with 4 fuses:
• Unscrew the fuse cap with a screwdriver.
• Remove both fuse cap and fuse.
• Insert the defined fuse into the sleeve, then use the screw
-
driver to screw the sleeve back in as far as it will go.
Fuse Protection
Type
F1
Display
T 0.25 A
F2
Coarse feed
T 1.6 A
F3
Processor board supply
T 1.0 A
F4
Peltier element
T 6.25 A
Fig. 56
Turn the instrument off and disconnect the power plug before replacing the fuses.
Use only the fuse types specified in Chapter "
3. Technical Data
".
Failure to comply with this can cause severe damage to the instrument and its surroundings
and a risk of fatal injury!