39
Leica CM1520 – Cryostat
7. Daily Use of the Instrument
10
9
8
6
7
Assembling the anti-roll guide system/replacing the anti-roll plate
1. Insert the glass into the interchangeable
frame and tighten it evenly using the knurled
screws (
7
).
2. Insert the shaft (
8
) of the metal frame for ex-
changeable glass inserts from above into the
bore of the swinging arm (
6
) in such a way that
the pin rests in the notch.
3. Push the white plastic plate (
9
) from below
onto the shaft (
8
).
4. Screw the knurled nut (
10
) from below onto
the shaft (
8
).
7.3.2 Blade holder CE-TC
The blade holder CE-TC is designed only for dis-
posable carbide metal blades (tungsten carbide
- TC65).
The handling of this holder is identical to that of
the blade holder CE.
Fig. 37
Fig. 38
All 4 lengths of the anti-roll guide
glass can be used.