Safety Notes

2  Safety Notes

2.1  Intended use

•  The CaptiView module is an optical instrument you can use to 

superimpose data or images. These overlays can be viewed 
superimposed over the microscopic image or on their own 
against a black background. Two users can observe the object 
under the microscope simultaneously at the optical outputs.  
In addition, you can connect a photo or video camera. 

X

X

Only use the connectors and cables included in the scope of 
delivery in order to connect the CaptiView. Use these connectors 
and cables when connecting it to an external device providing 
data as well.

•  The CaptiView can only be adapted to the Leica M530 optics 

carrier.

2.2  Non-intended use

•  Only use the CaptiView as described in this user manual. Failure 

to observe these instructions may result in property damage.

2.3  Information for the person 

responsible for the instrument

•  Ensure that only instructed users use the CaptiView.

•  Ensure that this User Manual is always available at the place 

where the instrument is used.

•  Carry out regular inspections to make sure the users are 

complying with safety requirements.

•  Brief the users thoroughly and explain the meaning of the 

warning signs and warning messages to them.

•  Allocate responsibilities for preparing the instrument for work, 

for operating it and for maintaining it. Monitor compliance with 
these.

•  Only use the CaptiView in proper condition.

•  Inform your Leica Microsystems representative or 

Leica Microsystems (Schweiz) AG, Medical Division,  
9435 Heerbrugg, Switzerland, immediately if you detect a 
product defect that could potentially cause injury or harm.

•  Ensure that the usual precautionary measures for 

electromagnetic and other radiation are observed.

•  Do not make any changes to this instrument.

•  Connected units and all the configurations for the analog and 

digital interfaces (signal inputs and outputs) have to be 
certified in accordance with IEC 60601-1/EN 60601-1 for medical 
devices. It is the responsibility of the user to ensure that the 
limits of the above-mentioned standards are not exceeded. If in 
doubt, please contact the appropriate technical department or 
the representative responsible.

•  The complete system does not conform to CE if the instrument is 

operated together with components that do not have 
CE certification.

2.4  Directions for the user of the 

instrument

•  Follow the instructions given by your employer regarding the 

organization of work and safety at work.

•  Follow the instructions given in this manual.

•  The electrical system at the installation location must conform 

to IEC, CEC and NEC specifications as applicable. 

•  Do not sterilize the CaptiView.

•  Observe the applicable statutory and country-specific 

regulations for accident prevention and environmental 
protection.

•  Check the function of the CaptiView before surgery.

2.5  Dangers of use

CAUTION
Danger of damage to the instrument due to improper 
storage!

X

X

Protect the instrument from direct sunlight and heat.

Summary of Contents for CaptiView

Page 1: ...MEDICAL DIVISION User Manual 10 746 768 Version 01 CaptiView From Eye to Insight...

Page 2: ...l Division Max Schmidheiny Strasse 201 CH 9435 Heerbrugg Phone 41 71 726 3333 Fax 41 71 726 3334 Legal disclaimer All of the technical data is subject to change without notice The information provided...

Page 3: ...6 5 2 Operating the CaptiView on the control unit of the Leica M530 OH6 OHX 6 6 Settings for CaptiView 10 6 1 Application 1 FL800 10 6 2 Application 2 IGS 11 6 3 Application 3 Common 12 6 4 Applicatio...

Page 4: ...located on the identification label on the CaptiView Enter this data in your user manual and always refer to it when you contact us or the service workshop regarding any questions you may have Model...

Page 5: ...Only use the CaptiView in proper condition Inform your Leica Microsystems representative or Leica Microsystems Schweiz AG Medical Division 9435 Heerbrugg Switzerland immediately if you detect a produc...

Page 6: ...746 768 Version 01 Design and function 3 Design and function 3 1 CaptiView controls 1 Emergency stop switch 3 2 Design and function of the handles for Leica M530 OH6 OHX 1 1 Joystick for menu navigati...

Page 7: ...t sockets 1 to 4 are assigned only with the following connections 1 XGA 3 IGS 2 XGA 2 FL800 3 XGA 1 General input e g Endoscopic camera 4 XGA Out To the CaptiView 4 1 2 Leica OHX LAN NAV USB ADF XGA I...

Page 8: ...IGS workstation Menu scroll down The keys of the handles can be configured separately on the left and right handle The joystick 2 can be configured for the X Y movement of the microscope or for menu c...

Page 9: ...ightness Decreases the brightness of the superimposed data 5 2 3 User settings You can configure the following settings for the user in the user settings Default brightness for image of overlay and no...

Page 10: ...lay Data Image Injection Channel It can be selected in which eye the image shall be injected Left Right Left Right Show Microscope Data Microscope data can be injected The injected data are Working di...

Page 11: ...guidance You do not need to configure any settings on the control unit 5 2 5 Safety function of the CaptiView If because of a malfunction the shutters cannot be controlled or the display switches itse...

Page 12: ...n of the XGA assignment 6 1 Application 1 FL800 FL800 Input To prepare the Leica operating microscope to display the signal of a Leica FL800 system for a specified user in the CaptiView User DIC setti...

Page 13: ...n 2 IGS IGS Input To prepare the Leica operating microscope to display the signal of an IGS system for a specified user in the CaptiView User DIC settings DIC In 1 No device DIC In 2 No device DIC In...

Page 14: ...on Common Input To prepare the Leica operating microscope to display the signal of e g an endoscope for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 No device DIC In 3...

Page 15: ...re the Leica operating microscope to display the signal of an IGS system and or a Leica FL800 for a specified user in the CaptiView User DIC settings DIC In 1 No device DIC In 2 FL800 DIC In 3 IGS Def...

Page 16: ...perating microscope to display the signal of a Leica FL800 and or e g an endoscope for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 FL800 DIC In 3 No device Default DIC...

Page 17: ...signal of an IGS system 1st priority and or e g an endoscope 2nd priority for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 No device DIC In 3 IGS Default DIC In DIC In...

Page 18: ...microscope to display the signal of an IGS system 2nd priority and or e g an endoscope 1st priority for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 No device DIC In 3...

Page 19: ...display the signal of a Leica FL800 1st priority an IGS system 2nd priority and or e g an endoscope 3rd priority for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 FL800...

Page 20: ...display the signal of a Leica FL800 1st priority an IGS system 3rd priority and or e g an endoscope 2nd priority for a specified user in the CaptiView User DIC settings DIC In 1 Common DIC In 2 FL800...

Page 21: ...To disinfect the CaptiView use compounds from the surface disinfectant group based on the following active ingredients aldehydes alcohols quaternary ammonium compounds Due to potential damage to the m...

Page 22: ...t a white balance for the video camera endoscope connected for the overlay X X If the external data source is a video camera carry out a white balance Cause 3 Defective converter The converter process...

Page 23: ...s remain closed Fault remedy X X Check the IGS workstation If the IGS workstation does not function correctly X X Press the switch for manually switching off the display and opening all shutters 18 re...

Page 24: ...ata Weight CaptiView 1 2 kg 9 4 Ambient conditions Transport 40 C to 70 C 40 F to 158 F Storage 10 C to 55 C 14 F to 131 F to 75 relative humidity Mode 10 C to 40 C 50 F to 104 F to 75 relative humidi...

Page 25: ...23 CaptiView Ref 10 746 768 Version 01...

Page 26: ...24 CaptiView Ref 10 746 768 Version 01 Specifications...

Page 27: ...25 CaptiView Ref 10 746 768 Version 01 Specifications...

Page 28: ...quality assurance and environmental management Leica Microsystems an international company with a strong network of worldwide customer services Active worldwide Tel Fax USA Buffalo Grove Illinois 1 80...

Reviews: