What To Do If...?
Fault:
The superimposed image is out of focus
Cause:
Uncorrected refractive error.
Fault remedy:
X
X
Correct the refractive error at the eyepieces (see the user
manual of your microscope).
8.2 Troubleshooting the shutters
Fault:
No microscope image when overlay is deactivated
Cause 1: The CaptiView receives incorrect commands from the IGS
workstation. The shutters remain closed.
Fault remedy:
X
X
Check the IGS workstation.
If the IGS workstation does not function correctly:
X
X
Press the switch for manually switching off the display and
opening all shutters (18) (refer to Chapter "CaptiView
controls", page CaptiView) and contact your IGS partner.
Cause 2: Defect in the CaptiView. The shutters remain closed.
Fault remedy:
X
X
Check the IGS workstation.
If the IGS workstation does not function correctly:
X
X
Press the switch for manually switching off the display and
opening all shutters (18) (refer to Chapter "CaptiView controls",
page CaptiView) and contact your Leica representative
Fault:
Shutters do not close/open as desired
Cause:
DIC settings set incorrectly.
The position of the shutters saved in the "DIC Settings"
menu is incorrect.
Fault remedy:
X
X
Check the saved positions of the shutters in the "DIC" menu of
the control unit (see page CaptiView) and set them as
required.
8.3 Troubleshooting the display
Fault:
The display shows a signal error
Cause 1: The display switches on by itself and/or shows an
incorrect video signal.
Fault remedy:
X
X
Press the switch for manually switching off the display and
opening all shutters (18) (refer to Chapter "CaptiView
controls", page CaptiView). Contact your Leica Microsystems
representative.