66
Operating Manual V 1.0 – 07/2006
7.1.3 Smart Clean
The smart clean prepares the ASP300 S for transport. During a smart clean,
a series of air cleaning steps will be applied which will remove any excess
liquid (condensate) from all internal tubing.
Prior to starting a smart clean, a complete retort clean program must be run.
To start a smart clean, press the
SMART CLEAN
button on the
MORE MENU
FUNCTIONS
screen and press
OK
to confirm.
The smart clean itself takes approx. 15 minutes and consists of 3 phases:
Phase 1:
Prior to starting the smart clean make sure all reagent containers are
correctly inserted and connected.
All internal tubing is cleaned with pressurized air to force any remaining
reagents back into the reagent bottles.
Phase 2:
Before phase 2 starts, the operator is prompted to remove all reagent
bottles from the instrument.
All internal tubing is again cleaned with pressurized air.
Phase 3
Before phase 3 starts, the operator is prompted to insert an empty bottle
into station 11.
The tubing is cleaned with pressurized air for the third time - remaining rests
of reagents, if any, are forced back into station 11.
Empty and clean all reagent bottles thoroughly before reinsert-
ing them into the instrument to make sure there will be no re-
agent spills from the bottles that could cause damage during
transport.
→
7.
Cleaning and Maintenance