35
Leica ASP300 S
Setting the altitude of the installation location (in meters above sea level)
This is important as it affects the actual pressure in
the retort.
Press
SITE ELEVATION (metres)
, enter the value in
the number field and press
OK
.
This parameter (in meters) must be entered to ensure
that the ASP300 S makes the appropriate corrections
when calculating the proper pressure or vacuum.
→
Updating software:
New versions of the ASP300 S operating software are provided on diskette.
Such upgrades generally contain two to three diskettes.
Do not perform a software upgrade if you do not have all of the
diskettes in the upgrade package.
Read the instructions in the upgrade pack with care, as they
may contain additional version-specific information as well as
new or modified procedures, parameters and data that must be
loaded or modified to ensure that the new software version
functions correctly.
1.
Make a backup copy of the instrument’s current data. You will need an
empty, formatted diskette for this purpose.
2.
Insert "Disk 1" (or similar) of the software upgrade into the diskette drive.
3.
Press
UPDATE SOFTWARE
, dismiss the safety queries with
YES
/
OK
and
follow the onscreen instructions.
Insert "Disk 2" into the diskette drive when prompted. Continue with the
remaining disks (if any).
4.
When the update process is complete, the instrument’s software will
reboot – the screen will switch off and restart as it would when switch-
ing the instrument off and back on.
↓
↓
5.
Operation