72
Check & Adjust
If the error is bigger than the predefined limit, the procedure has to be
repeated. The tilting axis measurements of the current run are then
rejected and not averaged with the results from previous runs.
6.
TOOLS T-Axis Adjustment Accuracy
<No.of Meas:>
Shows the number of runs executed. One run consists of
a measurement in face I and face II.
<
σ
a T-axis:>
shows the standard deviation of the determined tilting axis
error. The standard deviation can be calculated from the second run
onwards.
It is recommended to measure at least two runs.
7.
if more runs have to be added. Continue with step 3.
OR
to accept the measurements and to proceed to
TOOLS T-Axis
Adjustment Result
. No more runs can be added later.
Step
Description