The ATK HydraPro is designed to be used in harsh environments, including
water and dirty conditions. If subjected to these environments, the knee
should be cleaned immediately to prevent premature wear and tear. To clean
the knee, submerge it in clean warm freshwater or place knee under warm
running water. Dry after.
The ATK HydraPro (HYD-PA-02) and the ATK HydraPro with Stance Flexion
(HYD-SF-02) can be used in freshwater and saltwater, including swimming
pools. If the knee is exposed to non-freshwater, rinse thoroughly in clean
warm freshwater immediately after exposure to dissolve salts and limit
unnecessary exposure to chemicals.
Light corrosion will not affect function. In the event of heavy corrosion, con
tact
your prosthetist as it could affect function or safety.
PROSTHETIST
Maintenance:
The joint must be inspected, and repaired if necessary, at least every 6
months. Inspect: alignment, screw connections, suitability of user (weight,
degree of mobility), loss of lubricant, for damage, for soiling of the bush
-
ings, extension stop, visual inspection of proximal adaptor, other damage.
Maintenance must be carried out by prosthetist. Visual inspection annually
is recommended. Check for visual defects that may affect proper function.
Maintenance instructions: do not disassemble the knee joint. If there is
an issue with the product, send the knee joint in for service. LegWorks
recommends readjusting the knee joint’s settings once the user has gotten
used to the prosthesis.
Care:
Clean with soft cloth and warm freshwater. Do not use aggressive cleaning
agents, or compressed air.
Light corrosion will not affect function. In the event of heavy corrosion,
contact LegWorks as it could affect function or safety.