6300-M3 : TYPICAL USE
STEP 0: PREP AIRCRAFT
0.1
Connect adapters and hoses to the aircraft, tape-up all unused ports on the aircraft.
0.2 Leave hose ends near the 6300-M3 but DO NOT connect to 6300-M3 yet.
0.3 Set the BARO adjust on the aircraft to 29.92 or 1013 or Standard.
STEP 1: SELF TEST
1.1 Turn ON 6300-M3 and check / adjust leak timers in timed-leak screen (put both Pitot and Static in Leak). ).
After checking / adjusting leak-timers, press CANCEL to return to Main Screen.
1.2 Open at least one Pitot and one Static port on 6300-M3 to ambient. Close Vent valves.
1.3 Run SELF TEST (NEXT SCREEN twice; GO).
STEP 2: SET GROUND
2.1
With ports still open SET GROUND (F-7 ; GO)
STEP 3: LOW-LEVEL LEAK CHECK
3.1
Connect hoses from aircraft to 6300-M3. Check Yellow and Green ends match.
3.2
GO TO GROUND (F-8 ; GO)
3.3
Pitot Leak check:
a. Airspeed Target = 200 knots; GO
b. Setup Pitot Mode = Leak (“L” flashing) wait for Pitot to stabilize at 200 knots; GO
c. If Pitot cannot achieve 200 knots or if Pitot Leak exceeds 50 kts/min. (warning message ) then GO TO
GROUND, Go to MEASURE on BOTH Pitot and Static, then find the leak. Go back to Step 3.2
d. If Pitot leaks down to below 100 knots and leak has not yet been found then CANCEL out of Leak
Screen, allow Pitot to “recharge” to 200 knots, then change Pitot back into Leak Mode and continue
searching for the leak.
e. When Pitot is Leak tight, CANCEL out of Leak Screen and proceed to step 3.4
3.4
Static Leak check:
a. Airspeed at 200 knots; Altitude Target = 3000 feet above Ground; GO
b. Setup Static Mode = Leak (“L” flashing) wait for Static to stabilize at target; GO.
c. If Static cannot achieve target or if Static Leak exceeds 2000 ft/min. ( warning message) then GO TO
GROUND, Go to MEASURE on BOTH Pitot and Static, then find the leak. GO TO GROUND, then go
back to Step 3.4
d. If Static leaks down more than 1000 feet and leak has not yet been found then CANCEL out of Leak
Screen, allow Static to “recharge” to target, then change Static back into Leak Mode and continue
searching for the leak.
e. When Static is Leak tight, CANCEL out of Leak Screen and proceed to step 4.
STEP 4: CHECKOUT AIRCRAFT INSTRUMENTS
4.1 Run Profile (F -4) or execute your Test Procedure to checkout cockpit instruments.
4.2 GO TO GROUND at the end of Profile or Test Procedure.
4.3 Turn off 6300-M3, then disconnect hoses.