29
3.3.1 TIMED-LEAK SCREEN
When both Pitot and Static outputs are simultaneously put into Leak mode, then the leak
screen displays the leak values in a timed-leak format as shown below. The sample screen
shown below assumes the following:
a. Units are set to knots and feet
b. Leak timers are set for 1, 2 and 3 minutes
c. The leak-check was started at 100 knots and 15000 feet
d. The elapsed leak time is 3 minutes and 5 seconds
98.8 knots -27 ft/min. 14903 feet
1m: 99.5 / -0.5 14965 / -35
2m: 99.1 / -0.9 14932 / -68
3m: 98.8 / -1.2 3:05 14905 / -95
The first line shows the current Actual values. The second line shows the Actual values and
total leak at the end of 1 minute. The third line shows the Actual values and total leak at the
end of 2 minutes. The fourth line shows the Actual values and total leak at the of 3 minutes.
The fourth line also shows the elapsed leak time in the “mm:ss” format, in the center of the
line.
When both Static and Pitot are in leak mode and the you first enter the Leak screen, it looks
like the sample screen below. Since the elapsed time is at 0:00, only the current Actual
values are displayed, and an arrow points to the first leak-timer setting, which is shown to
be 1 minute. The arrow indicates that the first leak-timer value may be changed at this
point, if necessary.
Changing leak-timer values:
The leak-timer value can be changed by using the Up-Arrow and Down-Arrow keys. If
neither of these keys is pressed before the elapsed time reaches 10 seconds, then
all three
leak-timer values are maintained at their current setting and the arrow is cleared from the
display. The 1
st
leak-timer value can be adjusted between 1 and 18 minutes.
100.0 knots 0 ft/min. 15000 feet
1m:
2m:
3m: 0:00