ID Function
Unit
Command
68 Offset source X for set point: 0 = normal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt second
[–]
OUT_MODE_04_X
Note (ID 66 and 68): If X = 3, the commands ID 66 and ID 68 cannot be executed in some constant temperature control
devices until an external temperature specification has been received (via the command ID 15).
Table 24: Rights
ID Function
Unit
Command
62 Keyboard Master (equivalent to "KEY"): 0 = unlock / 1 = lock
[–]
OUT_MODE_00_X
64 Keyboard remote control unit (command): 0 = unlock / 1 = lock
[–]
OUT_MODE_03_X
Table 25: Status
ID Function
Unit
Command
74 Switch the device on / off (standby)
[–]
START / STOP
Table 26: Programmer
ID Function
Unit
Command
76 Select the program for the subsequent commands (X = 1 – 5). The
default program is 5 when the constant temperature equipment is
switched on.
[–]
RMP_SELECT_X
78 Start programmer
[–]
RMP_START
79 Pause programmer
[–]
RMP_PAUSE
80 Continue programmer (after pause)
[–]
RMP_CONT
81 End programmer
[–]
RMP_STOP