Example with set point transfer of 30.5 °C to the constant temperature
equipment.
PC / control station
Constant temperature equipment
"OUT_SP_00_30.5"CRLF
è
ç
"OK"CRLF
Example
The interface operates with 1 stop bit, 8 data bits and no parity bits.
4 transmission speeds are available for selection: 2400, 4800, 9600 or
19200 baud. 9600 baud is the factory preset.
RS 485 commands are always preceded by the device address; up
to 128 device addresses are available. Device addresses always contain
three digits and start with "A": "A000_..." – "A127_...".
Commands from an external source must always end with CR. The
response from the constant temperature equipment always ends with
CR.
Example with set point transfer of 30.5 °C to the constant temperature
equipment. In this example, the address 15 is used.
PC / control station
Constant temperature equipment
"A015_OUT_SP_00_30.5"CR
è
ç
"A015_OK"CR
Example
7.2.2 Read commands
The interface module recognizes the following read commands, which you
can use to retrieve operating data of the constant temperature equipment.
Table 5: Temperature
ID Function
Unit, resolu-
tion
Command
2 Temperature set point
[°C]
IN_SP_00
3 Bath temperature (outflow temperature)
[°C], 0.01 °C IN_PV_00
4 Bath temperature (outflow temperature)
[°C],
0.001 °C
IN_PV_10
5 Controlled temperature (internal / external Pt / external analog / external
serial)
[°C]
IN_PV_01
7 External temperature T
E
(Pt)
[°C]
IN_PV_03
8 External temperature T
E
(analog input)
[°C]
IN_PV_04
14 External temperature T
E
(Pt)
[°C],
0.001 °C
IN_PV_13
RS 485 protocol
V1
Interface module LRZ 913
19 / 39