The tank is ventilated through the tank lid. Once the device is filled, do
not tip it and never set it upside down!
Empty the device before transporting it.
1.
Slide and remove the cover over the filler nozzle towards the back of
the device.
2.
Open the tank lid by turning it in a counterclockwise direction.
3.
Fill the tank with heat transfer liquid. Use a standard spray bottle or a
funnel to fill it.
4.
Put the tank lid on and close it by turning it in a clockwise direction.
5.
Slide in the cover over the filler nozzle.
WARNING!
Contact with hot or cold heat transfer liquid
Scalding, cold burns
Bring the heat transfer liquid to room temperature before
draining.
Take the following information into account:
Empty the device when decommissioning it or if there is danger of frost!
The pump connections on the device are self-closing. This ensures that liquid
does not flow out of the device.
1.
The device is switched off.
2.
In order to empty the device, stick the hose nozzle into a pump con-
nection and hold the end of the hose over a suitable container.
Liquid will immediately flow out of the device.
The tank and the hydraulic line in the device will empty out parti-
ally.
3.
Plug the hose nozzle into the other pump connection.
The tank and the hydraulic lines in the device will empty out com-
pletely.
Filling the tank
Fig. 9: Tank lid
Empty the tank
18 / 41
LOOP
V3
Summary of Contents for LOOP 100
Page 21: ...5 2 LOOP menu structure Fig 10 Menu structure V3 LOOP 21 41...
Page 42: ......
Page 43: ......