1.7
Heat transfer liquid
The device is designed exclusively for non-combustible heat transfer liq-
uids according to class I as per DIN 12876-1.
The use of heat transfer liquids poses a risk of injury from high and low
temperatures if certain upper or lower temperature thresholds are
exceeded or the external application is broken and a reaction is initiated
with the heat transfer liquid.
All possible risks in handling the heat transfer liquid are specified in the
safety data sheet for the liquid together with corresponding safety
measures. The safety datasheet must therefore be observed in order to
use the device as intended.
1.8
Application area
The device may only be used in the following areas.
Commercial sector
Use inside buildings
Height up to 2,000 m
Ambient temperature of 5 to 40 °C
Maximum relative humidity 80 % at temperatures up to 31 °C, linearly
decreasing until 50 % relative humidity at 40 °C
Fluctuations of the mains voltage up to +/-10 % of the nominal voltage
Transient electrical surge up to the values of the surge category II
Sporadic electric surges, that occur in the mains power supply
Pollution degree 2
1.9
Personnel qualification
Before operating the device, the operating personnel must read the oper-
ating manual. The operator must have understood the operating manual.
1.10
Structure of warnings
A warning of "dangerous" indicates an immediately dangerous situation.
If this warning is not observed, then death or severe, irreversible injury
could occur.
DANGER!
Type and source
Consequences of not following instructions
Measure 1
Measure...