15.3
Setpoint offset operating mode (Diff.set/actual)
It is possible to apply an offset value to the temperature, which is provided by an external temperature sensor and to
process it as the set value.
The bath temperature can therefore be operated, for example, -15 °C below the temperature of a reactor measured
by the external temperature sensor.
-
Access to the main menu level is obtained by pressing the enter key
.
-
Selection and confirmation of
Setup
Control
Setpoint offset
.
The adjacent menu window appears.
As well as
Diff.set/actual value
the currently set offset value is displayed.
The adjacent menu window appears on selecting the menu point
Offset source
.
-
Select and confirm setpoint source with or and .
-
The setpoint offset is deactivated with "off".
-
Select and confirm
Diff.set/actual value
with or and .
The adjacent menu window appears. The minimum and maximum possi-
ble offset values and the current offset value are displayed.
-
Change the value with or .
-
Single figures can be selected by pressing or .
-
By pressing
(+/-) the arithmetic sign can be changed.
-
Confirm your choice with the enter key .
-
By pressing
(ESC) you are returned to the menu level without any change.
Summary of Contents for ECO GOLD
Page 1: ...Operation manual ECO Gold ...
Page 2: ......
Page 4: ......
Page 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Page 82: ...82 137 ECO Gold 10 2019 i ...
Page 83: ...10 2019 i ECO Gold 83 137 ...
Page 84: ...84 137 ECO Gold 10 2019 i ...
Page 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Page 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Page 130: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......