58 / 137
ECO Gold
8.2.3
Software version
SW version
-
confirm with .
Under the menu point SW version the appropriate software versions are
displayed, depending on the device type and connected modules.
8.2.4
Displaying and changing the device type
Type
-
confirm with .
The device type without the suffix "G" (GOLD) is shown in the menu.
You can change the device type .
Note:
With a change of device type parameters are re-initialized and control parameters adapted by the user are lost!
Therefore, the type change has a three second delay on the key depression.
The overtemperature switch-off point T
max
is automatically adapted to the device type, i.e. with the ECO GOLD ther-
mostat with a stainless steel bath T
max
= 202 °C, for the ECO GOLD thermostat with transparent bath T
max
= 102 °C.
You must now manually re-enter T
max
(
7.4.1), because otherwise the device enters the error status (error message in
ECO GOLD "T max diff. Ctrl-Safety").
8.2.5
Displaying serial numbers
Serial numbers
-
confirm with .
Under the Serial numbers menu point the serial numbers of Control and
Safety are displayed. Provided they are available, the serial numbers of
connected modules are also displayed.
Summary of Contents for ECO GOLD
Page 1: ...Operation manual ECO Gold ...
Page 2: ......
Page 4: ......
Page 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Page 82: ...82 137 ECO Gold 10 2019 i ...
Page 83: ...10 2019 i ECO Gold 83 137 ...
Page 84: ...84 137 ECO Gold 10 2019 i ...
Page 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Page 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Page 130: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......